Thermal pulse specific heat

Measurement principle

Thermal pulse specific heat measurement is based on flat-top pulsed magnetic field technology.We use high efficiency heating source and heat conduction material, high sensitivity low temperature thermometer, and small specific heat measurement chip to achieve pulse field specific heat measurement. First, the sample is cooled to the lowest temperature to be measured. Then a small amount of heat ∆Q is applied to the sample by means of a pulse current during the flat-top magnetic field. The heat capacity of the sample can be calculated by measuring the temperature change ∆T of the sample rise with a thermometer.

   

(a)the structure of the sample coupled with the thermal bath, thermometer and heater experiment parameters; (b) system structure of sample specific heat measurement by coating method

Experiment parameters

Magnetic field: 0-64T

Sample temperature: 1.5-400K

Measure precision:  ~10μJ/K

Measuring range:100μJ/K ~100mJ/K

Measurement direction:parallel magnetic field, vertical magnetic field

Sample size:1mg~100mg

  